⺣ - רҵ·ĵϷƽ̨

ǰλãҳ > 外文文献翻译 压电传感器测定动脉脉搏波速度 - 百度文库

外文文献翻译 压电传感器测定动脉脉搏波速度 - 百度文库

  • 62 Ķ
  • 3
  • 2025/12/12 4:25:47

Piezoelectric sensor determination of arterial pulse wave velocity

Arterial pulse wave velocity (APWV) is a measure of the elasticity (or stiffness) of peripheral arterial blood vessels. The pulse referred to here will be the pressure pulse as opposed to the flow pulse measured by ultrasound Doppler.

The pressure pulse velocity varies over the range from about 12 m s?1 to

15 m s?1 in stiff peripheral arteries, whereas in normal arteries it has a velocity in the range of 7 to 9 m s?1.

The aim of this project was the development of a fast and easy to use system for the determination of peripheral arterial pulse wave velocity. The principle of the PWV measurement is based on simultaneous measurement of two pulse waves at two different positions, such as the radial artery at the wrist and the brachial artery just above the elbow. By determining the pulse transit time between these points and the distance measured between the two locations, pulse wave velocity may then be calculated. The pressure pulse detection is done by using two piezoelectric sensors which generate a measurable voltage at the output contacts if they are mechanically deformed. The deformation produced voltage is first amplified and filtered and then digitalized with a data acquisition card. The analysis of the data obtained from the sensors includes a filtering process, the calculation of the PWV with three different methods foot-to-foot, cross-correlation and peak-to-peakand the determination of the arterial pulse rate.

Extensive measurements with human test subjects were carried out to optimize the techniques of data acquisition and analysis. For example, it was found that the best procedure was to hold the sensors in place using elastic straps alone. The data analysis was upgraded with an additional software module, which deletes, in effect, outriders or invalid measurements. With the optimized system, a series involving eight test subjects ranging in age from 22 to 32 years was completed (all normotensive). The arterial pulse wave velocities determined covered a range from 6 m s?1to 12 m s?1, with an average standard deviation of less than 2.5 m s?1for individual results. These are slightly higher,but close to published APWV data. The results showed that reproducible results can be obtained with the existing PWV acquirement and analysis system.

The measurement of arterial pulse wave velocity (APWV) is one of the methods used to measure physiological changes in peripheral vascular disease. Others include pressure pulse contour, arterial elasticity, pulsatile flow, complex vascular impedance and cardiac work.There have been many investigations over the past 30 years or so to relate changes in age and progress of arterial disease, to vessel pathology and distensibility based on the propagation characteristics of the arterial pulse (Malindzak and Meredith 1970, McCormack 1981, Persson et al 2001, Ramsey 1995, Wilkinson 1998). They have only been partially successful from a clinical point of view, mainly due to the difficulty in controlling the effects of patient parameters

(autonomic system, patient movement, etc).Early on, the methods of determining APWV involved direct invasive measurement of the pulse pressure at two points along an arterial segment and measuring the time taken for the pulse to travel the length of the segment. Such investigations had to be limited to animals,usually dogs.

Malindzak and Meredith (1970) carried out a comparative study of the arterial pulse wave velocity on anaesthetized dogs by recording intra-luminal pressure measured at two sites (proximal and distal) along the abdominal aorta axis.Non-invasive methods have included flow Doppler(McCormack1981) and tissue Doppler(Persson et al 2001) sensors using human test subjects. The comparative study referred to above was required because the pulse velocity measure ment canbe sign if i cantly modified ,and differs from the true pulse velocity (TAPWV) due to the viscosity of the fluid, the presence of reflected components (from bifurcations, stenoses and peripheral vascular beds) and due to changes in the elastic composition of the arterial wall. All of these lead to alterations in the pulsatile wave characteristics, so that what is measured is an apparent pulse wave velocity (AAPWV). The true PWV is that velocity of the arterial pressure pulse at which pressure waves would propagate in an infinite homogeneous tube (with a length much greater than the wavelength).

To reliably predict the elastic parameters (stiffness) of the arterial wall, which are of fund ament al importance to the understanding of the mechanics of the arterial system, depends on getting an accurate measure of the true arterial pulse wave velocity. Modelling the peripheral arterial segment as a thin-walled isotropic, incompressible tube contain ingan incompressible,

inviscid fluid, then the PWV is related to the wall elasticity by the MoensCKortweg

equation (Moens 1878), c0=(Eh/2R) (1)

where h is the wall thickness, R is the lumen radius, E is the wall elasticity modulus and is

the fluid (blood) density.

The sensor technique used in this work involves the piezoelectric effect in polyvinylidene fluoride (PVDF), which produces an output voltage in response to mechanical pressure on the material.

Three methods of APWV determination are used: foot-to-foot APWV; peak-to-peak APWV and cross-correlation APWV. The FFAPWV and CCAPWV methods are less sensitive to pressure wave reflections at bifurcations, etc in the arterial tree, than the PPAPWV method. Mean values and standard deviations were computed for all three methods and compared.

Foot-to-foot APWV (FFAPWV). This is based on the velocity of the foot, or leading edge, of the pressure pulse wave. The arrival times of the foot of the pulse wave at two positions along the artery are recorded. If t is the difference in arrival times and s the distance between the two recording positions (proximal and distal), the FFAPWV is simply

FFAPWV = s/t.

Peak-to-peak APWV (PPAPWV). This is completely analogous to the

FFAPWV except that the points of observation are the two (proximal and distal) peaks of the pulse wave and

PPAPWV = s/t.

Apparent pulse wave velocity (AAPWV). The pressure wave may be represented as a Fourier series,

P(t) = P0 +?Pn sin(nt + n)

n?1kWhere P0 is the mean fluid pressure, n is the harmonic number, Pn is the amplitude of the nth harmonic and n is the phase angle of the nth harmonic.

The spatial rate of change of the phase for one harmonic based on two

simultaneous pressure measurements separated by a distance _s along an artery, is related to the apparent arterial pulse wave velocity (AAPWV) by the following equation,

AAPWVn = (s)n(f )(360o)/(x1? x2)

Where AAPWVn is the apparent pulse wave velocity for the nth harmonic, f is the heart rate,x1 is the phase angle for the proximal harmonic n and x2 is the phase angle for the distal harmonic n.

Cross-correlation PWV. If the arterial pulse at the proximal measurement position is represented by the pressure time series P(x1, t) and that at the distal position by P(x2, t) and

the cross-correlation coefficient is x1,x2( ), then will have a maximum value at some time lag.

The correlation function can be expressed as

x1,x2( ) = (1/T )?/2?/2P(x1, t)P(x2, t) dt.

The value of at which maximum correlation occurs represents the transit time (_t) of the pressure wave from position x1 to position x2 along the arterial segment. From the separation distance and transit time data the correlation arterial pulse wave velocity is

CCAPWV = (x2 ? x1)/ t.

In this work normal, young test subjects were used, and it has the primary objectives of optimizing the measurement procedures and establishing the statistical spread and mean values of the observed PWVs for a specific peripheral arterial segment. Based on this, it is planned to use the system in clinical trials involving patients with peripheral arterial disease (due to diabetes, hypertension, etc), pre-, during and post treatment (pharmaceutical or surgical). Analogue and digital circuitry

Analogue charge amplifier. Piezoelectric materials convert mechanical stress or strain into proportionate electrical energy, by producing a charge when subjected to mechanical stress. The charge is converted to a voltage by an operational amplifier connected as a current integrator, called a charge amplifier. The signal output of the amplifier is approximately ?30 mV. It is augmented by signal amplification.

Analogue signal amplification . This is done by use of an inverting amplifier. Because a dc signal appears at the output of the charge amplifier, dc offset removal is essential and is implemented in the inverting summing

The next phase of the analogue circuitry is a low pass filter to remove the 50 Hz noise interference.

Digital controlled data acquisition and analysis. A data acquisition board (DAQ) is required when the transducer signals need interfacing with a PC. The board contains 12 bit plus sign and a successive approximation and self-calibrating

analogue-to-digital (ADC) converter. The ADC incurs a systematic error known as the quantization error. It is due to limited resolution and with the analogue input limit set at ?5 to +5 V, the quantization error of the A/D converter used here was calculated to be 0.122 mV.

The data acquisition and analysis was done using Lab Viewa powerful instrumentation and analysis programming language for PCs.

Digital data acquisition program. The data acquisition circuit performs all the necessary operations for the data acquisition with Lab View. The functions of the

circuit initialize the data acquisition and read the data from the acquisition card. These data are stored for later use in the data analysis part of the program.

Lab View programs are called virtual instruments (VIs) because of their

appearance and operations are analogous to measuring instruments. A VI that is called within another VI is called a sub-VI and is analogous to a subroutine in text-based languages.

The data acquisition, analysis and presentation are comprised of three main procedures:

(1) Data acquisition card to interface the hardware to the PC.

(2) Data acquisition program to acquire and store data in a spreadsheet file.

(3) Data analysis to carry out digital signal processing, calculate PWV and present results.

PWV Calculation

1. PWV calculation using peak detection. To calculate PWV using peaks, the location of the peaks must first be determined, so that the transit time of the wave

  • ղ
  • Υٱ
  • Ȩ
ĵ10.00 Ԫ VIP
ߣ...

92ƪĵ

ĵ飺

Piezoelectric sensor determination of arterial pulse wave velocity Arterial pulse wave velocity (APWV) is a measure of the elasticity (or stiffness) of peripheral arterial blood vessels. The pulse referred to here will be the pressure pulse as opposed to the flow pulse measured by ultrasound Doppler. The pressure pulse velocity varies over the r

οͿͨغɸƺŰ棩
ƪ
ʱؼۣ10 Ԫ/ ԭ:20Ԫ
VIP
ؼۣ29 Ԫ/ ԭ:99Ԫ
0.3 Ԫ/ ÿ150
ȫվɸ
VIP
ؼۣ29 Ԫ/ ԭ:99Ԫ
0.3 Ԫ/ ÿ150
ȫվɸ
עĵпܡֻĿ¼ݲȫ֮ǰעѸ޷ػ⣬ϵЭ㴦
΢ţfanwen365 QQ370150219
Copyright © ⺣ All Rights Reserved. ICP16052595-3 վͼ ͷQQ370150219 䣺370150219@qq.com